Nova Biomedical Corporation ABG Replacement Pump Assembly Catalog Number: 37092 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ABG Replacement Pump Assembly Catalog Number: 37092
Brand
Nova Biomedical Corporation
Lot Codes / Batch Numbers
Devices shipped between March 30-31, 2004
Products Sold
Devices shipped between March 30-31, 2004
Nova Biomedical Corporation is recalling ABG Replacement Pump Assembly Catalog Number: 37092 due to Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Recommended Action
Per FDA guidance
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, IL, LA, MA, NJ, NY, OH, PA, TX
Page updated: Jan 10, 2026