Novasys Medical Inc Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA
Brand
Novasys Medical Inc
Lot Codes / Batch Numbers
Issue affects all lots in distribution.
Products Sold
Issue affects all lots in distribution.
Novasys Medical Inc is recalling Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA due to Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects. Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation
Recommended Action
Per FDA guidance
A Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures. The company also updated physician information brochures and patient/physician DVDs to include the information. This was shipped to physicians in August/September of 2009. Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777. Questions should be addressed to the firm at the same telephone number.