NRT Nordisk Rontgen Teknik A/S Birkegaardsvej 16 Hasselager Denmark Celex/GE Precision MPi, REF no.: 02980000 Is an all-digital multipurpose tilt-C X-ray system, intended for a multitude of diagnostic procedures, including: R&F, radiology, fluoroscopy, interventional procedures, vascular and non-vascular procedures, and specialized applications including angiographic studies. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Celex/GE Precision MPi, REF no.: 02980000 Is an all-digital multipurpose tilt-C X-ray system, intended for a multitude of diagnostic procedures, including: R&F, radiology, fluoroscopy, interventional procedures, vascular and non-vascular procedures, and specialized applications including angiographic studies.
Brand
NRT Nordisk Rontgen Teknik A/S Birkegaardsvej 16 Hasselager Denmark
Lot Codes / Batch Numbers
Serial Numbers: 507003, 507004, 507005, 510004, 510009, 510010, 513001, 513003, 513007, 518001, 518002, 518004, 518005, 518006, 518011, 518013, 518014, 518015, 518017, 518018, 518020, 522001, 522002, 522003, 522004, 522005, 522006, 522008, 522009, 522012, 522014, 522015, 522016, 522018, 522019, 536001, 536002, 536003, 536006 and 536007.
Products Sold
Serial Numbers: 507003, 507004, 507005, 510004, 510009, 510010, 513001, 513003, 513007, 518001, 518002, 518004, 518005, 518006, 518011, 518013, 518014, 518015, 518017, 518018, 518020, 522001, 522002, 522003, 522004, 522005, 522006, 522008, 522009, 522012, 522014, 522015, 522016, 522018, 522019, 536001, 536002, 536003, 536006 and 536007.
NRT Nordisk Rontgen Teknik A/S Birkegaardsvej 16 Hasselager Denmark is recalling Celex/GE Precision MPi, REF no.: 02980000 Is an all-digital multipurpose tilt-C X-ray system, inten due to The GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes o. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The GE Precision MPi is in violation with 21 CFR Part 1020.32 requiring a signal audible to the fluoroscopist to sound for each passage of 5 minutes of fluoroscopic irradiation time during an examination or procedure.
Recommended Action
Per FDA guidance
An Advisory Notice concerning GE Precision MPi, dated May 1, 2009, was issued to Healthcare Professionals. The notice discussed the problem, risk, recommended immediate action, future actio, and affected system and serial numbers. A GE Field Service Representative will visit each affected U.S. customer site and introduce a speaker connected to the generator touch screen (to which the audible tone is delivered) in the operator's room, to be placed on the wall in the x-ray room. For further information, customers were to contact their local GE field engineer.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026