Ohmeda Medical, A Div of Datex-Ohmeda Inc Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.
Brand
Ohmeda Medical, A Div of Datex-Ohmeda Inc
Lot Codes / Batch Numbers
HDGDyyyyy HDGEyyyyy HDGFyyyyy HDGGyyyyy where ''yyyyy'' represents any number less than 60000
Products Sold
HDGDyyyyy HDGEyyyyy HDGFyyyyy HDGGyyyyy where ''yyyyy'' represents any number less than 60000
Ohmeda Medical, A Div of Datex-Ohmeda Inc is recalling Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed. due to Heater vent door on neonatal incubator may malfunction.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Heater vent door on neonatal incubator may malfunction.
Recommended Action
Per FDA guidance
The recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026