Olympus Corporation of the Americas EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190
Brand
Olympus Corporation of the Americas
Lot Codes / Batch Numbers
UDI-DI: 04953170305276 Serial: 2269931, 2611873, 2613235, 2634251, 2634326, 2860721, 2961773 and 2962545
Products Sold
UDI-DI: 04953170305276 Serial: 2269931, 2611873, 2613235, 2634251, 2634326, 2860721, 2961773 and 2962545
Olympus Corporation of the Americas is recalling EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (includ due to Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repa. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
Recommended Action
Per FDA guidance
Olympus issued Urgent Medical Device Correction letter via email on 4/22/22. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide for a service loaner in an effort to not disrupt patient care. Please do not hesitate to contact @ 647-999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026