Olympus Corporation of the Americas Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general open surgical procedures when used in conjunction with the Gyrus ACMI G400 workstation Ref: 970010PC Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general open surgical procedures when used in conjunction with the Gyrus ACMI G400 workstation Ref: 970010PC
Brand
Olympus Corporation of the Americas
Lot Codes / Batch Numbers
Lot number KR121617
Products Sold
Lot number KR121617
Olympus Corporation of the Americas is recalling Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and due to Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reacti. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reaction within the patient and / or electrical continuity to patient causing tissue burn to critical structures.
Recommended Action
Per FDA guidance
Olympus issued Urgent Medical Device Recall Letter on June 9th 2021 via Federal Express two day mail. Letter states reason for recall, health risk and action to take: 1. Immediately cease any further use of any affected product you have, remove it from your inventory and quarantine it until it is shipped back to us. 2. Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization to return any affected product. at no charge to you. Olympus will issue a credit or replacement to your facility for your affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0395 and provide your contact information as indicated in the portal. In addition, if you have further distributed this product, please identify your customers, notify them at once of this product recall, and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact me directly at 484-896-5688 or at laura.storms@olympus.com for any additional information concerning this matter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026