Olympus Corporation of the Americas SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
Brand
Olympus Corporation of the Americas
Lot Codes / Batch Numbers
UDI: 00815686020101 Serial Numbers: 85330200430, 87896220519, 87899220524, 87379210921, 87835220505, 87832220502, 87767220420, 87777220415, 87827220427, 87736220412, 87891220517, 87864220509, 87738220412
Products Sold
UDI: 00815686020101 Serial Numbers: 85330200430, 87896220519, 87899220524, 87379210921, 87835220505, 87832220502, 87767220420, 87777220415, 87827220427, 87736220412, 87891220517, 87864220509, 87738220412
Olympus Corporation of the Americas is recalling SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D) due to A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to troubleshoot the device
Recommended Action
Per FDA guidance
Olympus issued URGENT - Medical Device Field Corrective Action letter on 12/13/22. Letter states reason for recall, health risk and action to take: 1. Notify your users about this issue and stop using the single-use bronchoscopes (models INS-7100 or INS-7130) with SYS-4000 system until cable is replaced. Your users can continue to use the SYS-4000 using a reusable bronchoscope, with its respective video processor, without affecting their electromagnetic navigation (EMN) procedural experience. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter the recall number 0418 Provide your contact information as indicated in the portal. An Olympus Field Service Representative will contact you to schedule a site visit to replace the DVI cable. 3. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 to report complaints. Contact (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA