Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM
Lot Codes / Batch Numbers
ALL serial numbers/ controller lots are currently in scope of this recall
Products Sold
ALL serial numbers/ controller lots are currently in scope of this recall
A medical device manufacturer is recalling Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, due to The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or d. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Recommended Action
Per FDA guidance
On November 14, 2022, the firm initiated their Press Release at https://investor.insulet.com/news-releases/news-release-details/insulet-issues-nationwide-voluntary-medical-device-correction and notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to monitor their Omnipod 5 Controller for the following issues: - Charging port or charging cable appears melted, deformed, or discolored - Controller experiences extreme overheating (uncomfortable to hold) or emits an odor while charging If customers experience any of the issues above, they should disconnect the charger from the power outlet and do NOT charge the Omnipod 5 Controller. Promptly contact Insulet s dedicated Customer Care team specifically trained to help you at 1-800-641-2049 to request a replacement device. If you have a compatible Android smartphone (please see omnipod.com/compatibility), you may want to switch to using the Omnipod 5 App on your smartphone. Customers were provided with additional precautions to take to continue use of the device.
Verify this information on the official source
This page aggregates publicly available data. Always confirm recall details directly with the U.S. Food and Drug Administration (FDA) before taking action.
View official FDA recall noticePage updated: Jan 10, 2026
Important Notice
This page displays information sourced from the U.S. Food and Drug Administration (FDA). IsMyProductRecalled.com aggregates public data for convenience and does not issue recalls. We are not affiliated with any government agency. Always verify recall information with the official source before taking action. This is not medical or legal advice.