Ormco Corporation AdvanSync is an orthodontic device intended for the orthodontic treatment of Class II malocclusions. The AdvanSync part numbers and lot numbers affected by this recall are as follows: Part Number: 600-4000 (Kits-include 10 Right 600-4000 and 10 Left Assemblies). Lot Numbers: 070990588, 070976869, 060971515, 060971514, 04094278D Part Number: 600-4040 (5-Pack, Left Assembly). Lot Numbers: 080901333, 080902356, 070981416, 060971299, 060976084 Part Number: 600-4041 (5-Pack, Right Assembl Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
AdvanSync is an orthodontic device intended for the orthodontic treatment of Class II malocclusions. The AdvanSync part numbers and lot numbers affected by this recall are as follows: Part Number: 600-4000 (Kits-include 10 Right 600-4000 and 10 Left Assemblies). Lot Numbers: 070990588, 070976869, 060971515, 060971514, 04094278D Part Number: 600-4040 (5-Pack, Left Assembly). Lot Numbers: 080901333, 080902356, 070981416, 060971299, 060976084 Part Number: 600-4041 (5-Pack, Right Assembl
Brand
Ormco Corporation
Lot Codes / Batch Numbers
070976869, 060971515, 060971514, 04094278D Part Number: 600-4040 (5-Pack, Left Assembly). Lot Numbers: 080901333, 080902356, 070981416, 060971299, 060976084 Part Number: 600-4041 (5-Pack, Right Assembly). Lot Numbers: 080902357, 080901335, 070981417, 060971304, 060976079
Products Sold
Part Number: 600-4000 (Kits-include 10 Right 600-4000 and 10 Left Assemblies). Lot Numbers: 070990588, 070976869, 060971515, 060971514, 04094278D Part Number: 600-4040 (5-Pack, Left Assembly). Lot Numbers: 080901333, 080902356, 070981416, 060971299, 060976084 Part Number: 600-4041 (5-Pack, Right Assembly). Lot Numbers: 080902357, 080901335, 070981417, 060971304, 060976079
Ormco Corporation is recalling AdvanSync is an orthodontic device intended for the orthodontic treatment of Class II malocclusions. due to Manufacturing Error: There is a potential for a braze joint failure in the upper telescopic rod assembly. Although an incident could occur in which t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Manufacturing Error: There is a potential for a braze joint failure in the upper telescopic rod assembly. Although an incident could occur in which the eyelet loses connection from the rod, there is no risk of swallowing or aspirating the appliance because it will still be attached to the patient's crown.
Recommended Action
Per FDA guidance
Ormco sent a Urgent Medical Device Recall letter dated October 9, 2009,( via USPS 1st class mail) to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are instructed to return the affected lot numbers of the 5-Pack Left and Right Telescopic Rod Assemblies (600-4040 and 600-4041) and the Left and Right Telescopic Rod Assemblies contained within the AdvanSync Kit (600-4000). The affected parts are identified by the items contained inside the wells titled, "Left Rods" and "Right Rods." The remaining items within the kit are not affected by this recall notification; therefore, there is no need to return the entire kit. Customers were instructed that any product which they have in stock will be replaced at no charge or credited to their account. Customers were instructed to contact Ormco Customer Care at (800) 854-1741 (ext. 1) directly to handle the arrangements of a quick return and replacement. The affected product should be returned to Ormco Corporation at the following address: Ormco Corporation, 1332 South Lone Hill Ave., Glendora, CA 91740. Customers were instructed to label their returned product "RECALLED PRODUCT Attention: Customer Returns". In addition, customers were asked to complete the enclosed Return Form and return it by fax at (909) 962-5605.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026