Ortho-Clinical Diagnostics, Inc. VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Data Release Versions (DRVs) 6215 through 6223. In vitro diagnostic -VITROS Chemistry Products Calibrator Kit 11 is used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of vancomycin (VANC). Product Code: 6801696 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Data Disk (ADD) for Data Release Versions (DRVs) 6215 through 6223. In vitro diagnostic -VITROS Chemistry Products Calibrator Kit 11 is used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of vancomycin (VANC). Product Code: 6801696
Brand
Ortho-Clinical Diagnostics, Inc.
Lot Codes / Batch Numbers
UDI-DI: 10758750006601 Lot Number: 1172 Exp. 20-OCT-2023
Products Sold
UDI-DI: 10758750006601 Lot Number: 1172 Exp. 20-OCT-2023
Ortho-Clinical Diagnostics, Inc. is recalling VITROS Chemistry Products Calibrator Kit 11 VANC Reagent Generation (GEN) 50 with affected Assay Dat due to Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Unavailable to calibrate VITROS Chemistry Products VANC Reagent, Generation (GEN) 50, results in delay of testing
Recommended Action
Per FDA guidance
QuidelOrtho (formerly Ortho Clinical Diagnostics) issued Urgent Product Correction Notification Letter on 4/14/23 to Distributors and End/Users. Letter states reason for recall, health risk and action to take: Prior to calibrating VITROS VANC Reagent GEN 50 using VITROS Calibrator Kit 11, Lot 1172, install ADD DRV 6224 or above on your VITROS System(s). " Complete the enclosed Confirmation of Receipt form no later than April 24, 2023. " Forward this notification if VITROS Calibrator Kit 11, Lot 1172 was distributed outside of your facility. " Save this notification with your user documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until ADD DRV 6224 or above has installed. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. Contact Information If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.