Ortho-Clinical Diagnostics VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests.
Brand
Ortho-Clinical Diagnostics
Lot Codes / Batch Numbers
01-5846, Exp. 31 May 2006, 01-6036 Exp 31 October 2006, 01-6239 Exp 31 January 2007
Products Sold
01-5846, Exp. 31 May 2006; 01-6036 Exp 31 October 2006; 01-6239 Exp 31 January 2007
Ortho-Clinical Diagnostics is recalling VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnosti due to Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
Recommended Action
Per FDA guidance
Letters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.