Orthofix, Inc Orthofix Spinal Implants; Pillar SA PEEK Spacer System Instrument Case. Case contents: Implant Insertion Instrument, Straight Tamp, Bone Awl Drill Guide, Straight Drill, U-Joint Driver with and without retention, Flex Shaft Driver, Trial Insertion Instrument, Drill Tips, Cover Plate Inserter, Cover Plate Driver, Cover Plate Holder, Ratcheting Handle, Torque Limiting Handle, Hex Driver, Distractors/Sizers, Sizers REF: 49-0021: Orthoflex Spinal Implants, 1211 Hamburg Tnpk, Suite 300, Wayne, NJ 0 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Orthofix Spinal Implants; Pillar SA PEEK Spacer System Instrument Case. Case contents: Implant Insertion Instrument, Straight Tamp, Bone Awl Drill Guide, Straight Drill, U-Joint Driver with and without retention, Flex Shaft Driver, Trial Insertion Instrument, Drill Tips, Cover Plate Inserter, Cover Plate Driver, Cover Plate Holder, Ratcheting Handle, Torque Limiting Handle, Hex Driver, Distractors/Sizers, Sizers REF: 49-0021: Orthoflex Spinal Implants, 1211 Hamburg Tnpk, Suite 300, Wayne, NJ 0
Brand
Orthofix, Inc
Lot Codes / Batch Numbers
Part number: 49-1100, Lot Number 17301-L125.
Products Sold
Part number: 49-1100, Lot Number 17301-L125.
Orthofix, Inc is recalling Orthofix Spinal Implants; Pillar SA PEEK Spacer System Instrument Case. Case contents: Implant Inser due to The Pillar SA Implant Insertion Instrument 49-1100 may experience resistance in turning the activation knob while releasing an implant after intra-ope. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Pillar SA Implant Insertion Instrument 49-1100 may experience resistance in turning the activation knob while releasing an implant after intra-operative positioning.
Recommended Action
Per FDA guidance
A notification letter dated November 19, 2009 was sent to customers via Federal Express. The letter described the recalled product (PILLAR SA Implant Insertion Instrument, 49-1100) and problem and actions for customers to take. The affected product, PILLAR SA lmplant lnsertion lnstrument, 49-1100 is in the PILLAR SA set (set part number 49-0021) which was shipped to customers. An enclosed attachment provide instructions on product removal and return. Customers were instructed to complete the enclosed Tracking and Verification Form and fax to Orthofix Spinal Implants at 1-973-406-2912. A replacement instrument will be provided as available. In the interim, the Implant Inserter (Part Number 49-1000) included in the PILLAR SA UPE set (set number 49-0020) that was provided earlier this year may be used as an alternative to the affected product. For questions regarding the removal and return of the recalled product to Orthofix Spinal Implants, please contact your Account Services Representative. Direct other questions to the Regulatory Affairs Department at 1-973-406-2812. Direct product questions to Marketing at 1-862-221-1335 or Product Development at 1-973-406-2824.