Orthosoft, Inc. dba Zimmer CAS NDI P7 Position Sensor, Stereotaxic Instrument Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
NDI P7 Position Sensor, Stereotaxic Instrument
Brand
Orthosoft, Inc. dba Zimmer CAS
Lot Codes / Batch Numbers
Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)
Products Sold
Part Number: NDI P7 Position Sensor Product Number 201.053 (a component of part number 521.035, Sesamoid Plasty Camera Arm Kit) Lot/Serial Number: 38 units affected: P7-00321, P7-00339, P7-00360, P7-00361, P7-00362, P7-00363, P7-00364, P7-00365, P7-00366, P7-00368, P7-00369, P7-00370, P7-00371, P7-00372, P7-00373, P7-00375, P7-00387, P7-00412, P7-00414, P7-00415, P7-00416, P7-00417, P7-00418, P7-00419, P7-00420, P7-00421, P7-00422, P7-00423, P7-00431, P7-00451, P7-00452, P7-00453, P7-00562, P7-00563, P7-00565, P7-00568, P7-00569, P7-00590 (The lot number and the serial number are the same)
Orthosoft, Inc. dba Zimmer CAS is recalling NDI P7 Position Sensor, Stereotaxic Instrument due to Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS work. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Zimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.
Recommended Action
Per FDA guidance
Zimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the Zimmer CAS customer service, 1-866-336-7846.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA