Ossur H F Grjothals 5 Reykjavik Iceland Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020
Brand
Ossur H F Grjothals 5 Reykjavik Iceland
Lot Codes / Batch Numbers
REF/UDI-DI: L-125000/05690967566142, L-621200/05690967596590, L-621000/05690967118334, L-621100/05690967155322, L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.
Products Sold
REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.
Ossur H F Grjothals 5 Reykjavik Iceland is recalling Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: due to A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.
Recommended Action
Per FDA guidance
On 5/12/23, recall notices were sent to customers informing them of the following: a. For affected lots in use, replace both lock body and attachment pin as soon as possible for users that have the following risk factors: - Trouble with socket fit due to volume changes in the residual limb. - Had lock replacement without having new attachment pin installed at the same time. - Have non-Ossur attachment pin in use. - Complaint about loose fit of the socket or locking issues. b. For other users, replace the affected lots of lock body and attachment pin during the next planned consultation, such as liner replacement. If a user suggests there is an issue with the locking mechanism, the lock and attachment pin should be inspected for wear, regardless of lot number. Replace the lock and attachment pin if there are signs of wear; instructions for inspection of wear are attached. c. INFORM applicable personnel handling the products about the added warnings in the Instructions for Use. Firm recommends that you reach out to your patients as soon as possible to inform them about the potential risk of the pin disengaging from the lock, leading to loss of suspension. d. Return unused affected devices. e. Pass this notice to all who need to be aware within your organization or to any other organizations/customers who received affected product. f. Maintain the notice for awareness. g. Complete and return the acknowledgement form to recalls@ossur.com If you have questions or concerns about this notification, please contact customer service. Tel: 1-800-233-6263.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, WY, PR
Page updated: Jan 10, 2026