Paltop Advanced Dental Solutions Ltd 5, Ha-Shita Caesarea Israel Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020
Brand
Paltop Advanced Dental Solutions Ltd 5, Ha-Shita Caesarea Israel
Lot Codes / Batch Numbers
UDI: 07290108695649, Lot Number WO-011320
Products Sold
UDI: 07290108695649; Lot Number WO-011320
Paltop Advanced Dental Solutions Ltd 5, Ha-Shita Caesarea Israel is recalling Paltop Dynamic Dental Implant Dia 3.25mm L 13.0 mm, Catalog Number 21-70020 due to Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw w. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.
Recommended Action
Per FDA guidance
Urgent: Medical Device Recall notification letters dated 11/18/19 were sent to customers. Please discontinue use. Attached is a Reply Communication form that should be used to let us know whether the product has been used or will be returned. Once we receive that Form, a Customer Service agent will reach out to you with return instructions and provide a label if necessary. Paltop Advanced Dental will replace the implants that are in stock.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, FL, GA, MD, MA, NJ, NY, SC, TX, VA, DC