Philips Healthcare Inc. CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. It is indicated for use by healthcare providers whenever there is a need for generation of a patient record and computation of drug dosage.
Brand
Philips Healthcare Inc.
Lot Codes / Batch Numbers
Serial Numbers: 4621A00101, 4621A00102, 4621A00103, 4621A00104, 4621A00105, 4621A00106, 4621A00107, 4621A00108, 4621A00109, 4621A00110, 4621A00111, 4621A00112, 4621A00113, 4621A00114, 4621A00115, 4621A00116, 4621A00117, 4621A00118, 4621A00119, 4621A00120, 4621A00121, 4621A00122, 4621A00123, 4621A00124, 4621A00125, 4621A00126, 4621A00127, 4621A00128, 4621A00129, 4621A00130, 4621A00131, 4621A00132, 4621A00133, 4621A00134, 4621A00135, 4621A00136, 4621A00137, 4621A00138, 4621A00139, 4621A00140, 4621A00141, 4621A00142, 4621A00143, 4621A00144, 4621A00145, 4621A00146, 4621A00147, 4621A00148, 4621A00149, 4621A00150, 4621A00151, 4621A00152, 4621A00153, 4621A00154, 4621A00155, 4621A00156, 4621A00157, 4621A00158, 4621A00159, 4621A00160, 4621A00161, 4621A00162, 4621A00163, 4621A00164, 4621A00165, 4621A00166, 4621A00167, 4621A00168, 4621A00169, 4621A00170, 4621A00171, 4621A00172, 4621A00193, 4621A00194, 4621A00195, 4621A00196, 4621A00197, 4621A00198, 4621A00199, 4621A00200, 4621A00201, 4621A00202, 4621A00203, 4621A00204, 4621A00205, 4621A00206, 4621A00207, 4621A00208, 4621A00209, 4621A00210, 4621A00211, 4621A00213, 4621A00214, 4621A00217, 4621A00218, 4621A00219, 4621A00220, 4621A00221, 4621A00222, 4621A00223, and 4621A00224.
Products Sold
Serial Numbers: 4621A00101, 4621A00102, 4621A00103, 4621A00104, 4621A00105, 4621A00106, 4621A00107, 4621A00108, 4621A00109, 4621A00110, 4621A00111, 4621A00112, 4621A00113, 4621A00114, 4621A00115, 4621A00116, 4621A00117, 4621A00118, 4621A00119, 4621A00120, 4621A00121, 4621A00122, 4621A00123, 4621A00124, 4621A00125, 4621A00126, 4621A00127, 4621A00128, 4621A00129, 4621A00130, 4621A00131, 4621A00132, 4621A00133, 4621A00134, 4621A00135, 4621A00136, 4621A00137, 4621A00138, 4621A00139, 4621A00140, 4621A00141, 4621A00142, 4621A00143, 4621A00144, 4621A00145, 4621A00146, 4621A00147, 4621A00148, 4621A00149, 4621A00150, 4621A00151, 4621A00152, 4621A00153, 4621A00154, 4621A00155, 4621A00156, 4621A00157, 4621A00158, 4621A00159, 4621A00160, 4621A00161, 4621A00162, 4621A00163, 4621A00164, 4621A00165, 4621A00166, 4621A00167, 4621A00168, 4621A00169, 4621A00170, 4621A00171, 4621A00172, 4621A00193, 4621A00194, 4621A00195, 4621A00196, 4621A00197, 4621A00198, 4621A00199, 4621A00200, 4621A00201, 4621A00202, 4621A00203, 4621A00204, 4621A00205, 4621A00206, 4621A00207, 4621A00208, 4621A00209, 4621A00210, 4621A00211, 4621A00213, 4621A00214, 4621A00217, 4621A00218, 4621A00219, 4621A00220, 4621A00221, 4621A00222, 4621A00223, and 4621A00224.
Philips Healthcare Inc. is recalling CareVue Chart Release C.O Part Number: 862246 Intended for use in data collection, storage, and man due to Software: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication with. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Software: A software defect under certain circumstances could lead to the unintended administration of a duplicate dose of a scheduled medication within a single 24 hour period
Recommended Action
Per FDA guidance
Philips Medical notified accounts by letter on 5/6/09 the Urgent - Medical Device Correction notification. While customers await the correction of their devices, customers are given the following instructions: Clinicians should not rely solely on the worklist within CareVue Chart to determine actionable orders for the patient until their software is upgraded. Clinical users should always use the medication administration record (MAR) and the administration record in determining the care their patients receive in addition to the critical care worklist. Philips is currently developing a correction that will be available at no additional charge. A Philips representative will notify customers when the software correction becomes available. If further assistance is needed, customers are to contact the Philips Customer Care Center at 1-800-722-9377.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CA, FL, MA, OH, OK, RI, VT, WV