Philips Medical Systems (Cleveland) Inc IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed
Brand
Philips Medical Systems (Cleveland) Inc
Lot Codes / Batch Numbers
IntelliSpace Portal DX/HX/EX: 96027, 96029, 96025, 96030, 96031, 85479, 79935, 98087, 85175, 96024, 106013, 450003, 950002, 950003, 35009, 35030, 870056, 550012, 85081, 97069, 45001985333, 35036, and 85382. IntelliSpace Portal DX/HX/EX Upgrade: 950100, 55003, and 81178. IntelliSpace Portal DX/HX/EX Demo: 85451 and 87290. IntelliSpace Portal IX: 50306, 40219, 881030
Products Sold
IntelliSpace Portal DX/HX/EX: 96027,96029,96025,96030,96031,85479,79935, 98087,85175,96024,106013,450003, 950002, 950003,35009,35030,870056, 550012, 85081, 97069, 45001985333, 35036, and 85382. IntelliSpace Portal DX/HX/EX Upgrade: 950100, 55003, and 81178. IntelliSpace Portal DX/HX/EX Demo: 85451 and 87290. IntelliSpace Portal IX: 50306, 40219, and 40368. IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030
Philips Medical Systems (Cleveland) Inc is recalling IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number due to Software defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size t. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Software defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size the Transcatheter heart valve (THV) to be implanted .
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
FL, IN, NC, VT
Page updated: Jan 10, 2026