Philips Medical Systems North America Co. Phillips Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work, excluding mammography. It is designed for radiographic examination of the standing or seated patient or the recumbent patient in combination with a mobile X-ray table (trolley). The intended use is from direct digital imaging using the built in flat panel detector and in addition for free exposures on radiographic cassettes. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, including specialist areas like trauma or pediatric work, excluding mammography. It is designed for radiographic examination of the standing or seated patient or the recumbent patient in combination with a mobile X-ray table (trolley). The intended use is from direct digital imaging using the built in flat panel detector and in addition for free exposures on radiographic cassettes.
Brand
Philips Medical Systems North America Co. Phillips
Lot Codes / Batch Numbers
Devices were identified as SITE numbers: 553491, 554498, 554944, 553270, 554358, 556089, 556275, 555457, 556080, 557203, 41446107, 41446304, 41443692, 41446097, 41445359, 41445529, 41585399, 41586528, 41796147, 41958550, 42252302, 42616819, and 42865052.
Products Sold
Devices were identified as SITE numbers: 553491, 554498, 554944, 553270, 554358, 556089, 556275, 555457, 556080, 557203, 41446107, 41446304, 41443692, 41446097, 41445359, 41445529, 41585399, 41586528, 41796147, 41958550, 42252302, 42616819, and 42865052.
Philips Medical Systems North America Co. Phillips is recalling Essenta DR is a digital multifunctional X-ray systems, suitable for all routine radiographic exams, due to When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be se. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
When a patient is selected in the patient list and examination tab pressed with background query running at the same time, the wrong patient may be selected.
Recommended Action
Per FDA guidance
Philip Medical System sent an "URGENT-Device Correction" letter dated August 13, 2008 to their customers. The letter describes the recalled product as Essenta DR Patient and Image Data Software (SW) release 1.0. The firm informs the customers that when patient is selected in patient list and examination tab pressed with background query running at the same time, there is a risk that the wrong patient is selected on the Essenta DR Computed Radiography System. Customers are advised to verify the selected patient's information before releasing an exposure while the firm works on upgrading the Software. All affected units will receive a software upgrade to correct the issue. Customers should contact Philips Customer Care Service at 800-772-9377, #5, #4, #1 and reference FCO 712000031 for any questions or support concerning the recall.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.