Philips Medical Systems North America Co. Phillips Philips BV Endura Mobile X-Ray system with the extended rotation option. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Philips BV Endura Mobile X-Ray system with the extended rotation option.
Brand
Philips Medical Systems North America Co. Phillips
Lot Codes / Batch Numbers
Devices are identified with Site Numbers: 548598, 550838, 551589, 551695, 552444, 552467, 552468, 552669, 553171, 553860, 553917, 553976, 553989, 554350, 554497, 555001, 555021, 555243, 555395, 555463, 555517, 556135, 556171, 556531, 556587, 556608, 556612, 556678, 556862, 556977, 556979, 557037, 557063, 557590, 558233, 558237, 558613, 558975, 559070, 559663, 559665, 41416327, 41443860, 41445126, 41445430, 41445445, 41445552, 41566246, 41585083, 41658558, 41765654, 41796621, 41905206, 41938039, 41955473, 42045315, 42162893, 42291865, 42326711, 42373518, 42382482, 42394051, 42608281, 42639034, 42824768, 42861875, 42869100, 42869163, 42880805, 42966787, and 42975222.
Products Sold
Devices are identified with Site Numbers: 548598, 550838, 551589, 551695, 552444, 552467, 552468, 552669, 553171, 553860, 553917, 553976, 553989, 554350, 554497, 555001, 555021, 555243, 555395, 555463, 555517, 556135, 556171, 556531, 556587, 556608, 556612, 556678, 556862, 556977, 556979, 557037, 557063, 557590, 558233, 558237, 558613, 558975, 559070, 559663, 559665, 41416327, 41443860, 41445126, 41445430, 41445445, 41445552, 41566246, 41585083, 41658558, 41765654, 41796621, 41905206, 41938039, 41955473, 42045315, 42162893, 42291865, 42326711, 42373518, 42382482, 42394051, 42608281, 42639034, 42824768, 42861875, 42869100, 42869163, 42880805, 42966787, and 42975222.
Philips Medical Systems North America Co. Phillips is recalling Philips BV Endura Mobile X-Ray system with the extended rotation option. due to The C-arm rotation brake does not always function properly, and may not hold the C-arm position if the C-arm is not positioned in its maximum rotation. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The C-arm rotation brake does not always function properly, and may not hold the C-arm position if the C-arm is not positioned in its maximum rotation and if it is accidentally bumped by medical staff or patients. Unwanted C-arm movement may cause the C-arm to collide with a patient. The brakes function properly if the C-arm is rotated to its maximum positions.
Recommended Action
Per FDA guidance
Philips Medical Systems issued an Urgent Product Correction letter dated September 30, 2008 to consignees. The letter identified the affected device, what the problem is and under what circumstances it can occur, and the actions that should be taken by the customer or user in order to prevent risks for patients or users. Philips will contacdt customers to implement Field Corrective Action (FCO) #71800026), which describes a mechanical brake modification which will be installed by Philips. Until the Field Corrective Action has been completed, users should ensure that the C-arm is placed in the maximum extended position, or ensure that the C-arm is not touched during critical procedures if the C-arm is not in a maximum position. The Philips Call Center can be contacted at 800-722-9377
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.