Philips Medical Systems Philips M8000-64100 Table Mount Shipped standard with all IntelliVue Patient Monitors MP40-70, Gas Modules Gl and G5 and certain options of the IntelliVue Patient Monitors MP5 MPST and MP20/30. Product shipped between 6/03/08 and 9/12/08 as follows: M1013A IntelliVue G1 Gas Module M1019A IntelliVue GS Gas Module M8001A option E22 Quick release mount for IntelliVue MP20 M8002A option E22 Quick release mount for IntelliVue MP30 M8003A IntelliVue MP40 M8004A IntelliVue MPSO M8005A IntelliVue MP6 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Philips M8000-64100 Table Mount Shipped standard with all IntelliVue Patient Monitors MP40-70, Gas Modules Gl and G5 and certain options of the IntelliVue Patient Monitors MP5 MPST and MP20/30. Product shipped between 6/03/08 and 9/12/08 as follows: M1013A IntelliVue G1 Gas Module M1019A IntelliVue GS Gas Module M8001A option E22 Quick release mount for IntelliVue MP20 M8002A option E22 Quick release mount for IntelliVue MP30 M8003A IntelliVue MP40 M8004A IntelliVue MPSO M8005A IntelliVue MP6
Brand
Philips Medical Systems
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Product shipped between 6/03/08 and 9/12/08 as follows: M1013A IntelliVue G1 Gas Module M1019A IntelliVue GS Gas Module M8001A option E22 Quick release mount for IntelliVue MP20 M8002A option E22 Quick release mount for IntelliVue MP30 M8003A IntelliVue MP40 M8004A IntelliVue MPSO M8005A IntelliVue MP60 M8007A IntelliVue MP70 M8105A option E22 IntelliVue MP5 Quick Release Mount M8105AT option E22 IntelliVue MPST Quick Release Mount M8040A option U02 U02 Table top mount kit M8040A option E22 Quick release kit for MP20MP30 A list of S/N's can be obtained fromthe recalling firm.
Philips Medical Systems is recalling Philips M8000-64100 Table Mount Shipped standard with all IntelliVue Patient Monitors MP40-70, Gas due to Table Mounts assembled with incorrect length hardware may fall. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Table Mounts assembled with incorrect length hardware may fall
Recommended Action
Per FDA guidance
Philips issued an "Urgent Medical Device Correction" notification to all potentially affected customers on 10/22/08 via UPS for the US customers. The letter mailed to customers outside the US by Philips' regulatory contacts in each country via a tracked method. The letter informs customers of the problem and gives instructions on how to identify affected table mounts. Field Service Engineers will be dispatched to each customer site to inspect each Table Mount. If customers detect any sign of a loose connection between the monitor and the mounting plate while they are awaiting the Field Service Engineer, they are instructed to make sure that the monitor is not in a position to fall. They are asked to disconnect the monitor from the table mount. For additional information, contact Philips Medical Systems at 1-978-687-1501.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.