Philips Medical Systems Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All four model numbers have the same user manual. The device is intended for monitoring, recording and alarming of multiple physiological parameters. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All four model numbers have the same user manual. The device is intended for monitoring, recording and alarming of multiple physiological parameters.
Brand
Philips Medical Systems
Lot Codes / Batch Numbers
SureSigns VS3, Model numbers 863071, 863072, 863073, 863074
Products Sold
SureSigns VS3; Model numbers 863071, 863072, 863073, 863074
Philips Medical Systems is recalling Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 86 due to Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureS. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureSigns VS3 Vital Signs Monitor was not in Interval non-invasive blood pressure (NBP) mode. These alarms are only enabled when interval NBP mode is selected. If when monitoring SpO2, the clinician does not notice when the sensor is no longer attached, the patient is not being monitored through Spo2.This could result in delay or lack of needed therapy. The labeling provided with the monitor did not fully describe this feature.
Recommended Action
Per FDA guidance
All affected customers will receive the FSN (Field Safety Notice )informing them about the labeling correction. Included with the FSN will be an addendum to the instructions for use. customers are asked to read the addendum and insert a copy into each of their copies of the VS3 instructions for use. Customers will also receive a Medical Device Correction Confirmation Card that they will be asked to fill out and fax back to their local Philips representative. Contact your local Philips support organization at 1-800-722-9377 in the United States or 1-800-323-2280 in Canada.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.