Philips Medical Systems Sales & Service Region No. America Trolley for Intera/NT MRI System Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Trolley for Intera/NT MRI System
Brand
Philips Medical Systems Sales & Service Region No. America
Lot Codes / Batch Numbers
The firm uses site numbers instead of serial numbers: Site Numbers: 105313, 104340, 105929, 84700, 103811, 103904, 504899, 102157, 103592, 85500, 87355, 86404, 103247, 87604, 505682, X1595, 101595, 87062, 103215, 86012, 103620, 101070, 104897, 105486, 83306, 86184, 100826, 84899, 100629, 87527, 84820, 100786, 86085, X1649, 504412, 504741, 87365, 102786, 104886, 103688, 86570, 102095, 103665, 100459, 103611, 102353, 104102, 100462, 103101, 103540, 85158, 103450, 102671, 102896, 102698, 83080, 103782.
Products Sold
The firm uses site numbers instead of serial numbers: Site Numbers: 105313, 104340, 105929, 84700, 103811, 103904, 504899, 102157, 103592, 85500, 87355, 86404, 103247, 87604, 505682, X1595, 101595, 87062, 103215, 86012, 103620, 101070, 104897, 105486, 83306, 86184, 100826, 84899, 100629, 87527, 84820, 100786, 86085, X1649, 504412, 504741, 87365, 102786, 104886, 103688, 86570, 102095, 103665, 100459, 103611, 102353, 104102, 100462, 103101, 103540, 85158, 103450, 102671, 102896, 102698, 83080, 103782.
Philips Medical Systems Sales & Service Region No. America is recalling Trolley for Intera/NT MRI System due to Patient and MRI tabletop may fall if the tabletop is not positioned correctly on the MRI trolley.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Patient and MRI tabletop may fall if the tabletop is not positioned correctly on the MRI trolley.
Recommended Action
Per FDA guidance
The firm sent a letter via certified mail to all customers on May 13, 2004 instructing their customers to check the position of the tabletop to ensure it is correctly aligned on the trolley. The firm is implementing a design modification beginning May 13, 2004. A field service engineer will contact the customers to arrange for installation of the retrofit for the trolley.
Verify this information on the official source