Philips North America, LLC Cardio Vascular-Allura Centron, system code 722400 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Cardio Vascular-Allura Centron, system code 722400
Brand
Philips North America, LLC
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Not distributed in US. Serial numbers 132 133 49 66 71 135 31 36 79 69 137 149 147 109 136 108 103 93 53 127 129 129 55 74 2 57 106 56 86 68 157 78 64 63 23 51 100 97 102 113 90 44 30 101 77 83 67 151 22 70 96 116 131 35 82 9 91 43 61 161 28 34 27 19 144 92 168 172 173 122 197 139 6 76 176 16 130 60 80 184 112 81 89 42 110 155 33
Philips North America, LLC is recalling Cardio Vascular-Allura Centron, system code 722400 due to A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Recommended Action
Per FDA guidance
On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.