Philips North America, LLC HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9.
Brand
Philips North America, LLC
Lot Codes / Batch Numbers
All units are affected by this recall
Products Sold
All units are affected by this recall
Philips North America, LLC is recalling HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M due to Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver ther. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.
Recommended Action
Per FDA guidance
On February 26, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that if the device is dropped or subjected to a severe mechanical shock, it may suffer internal damage even in the absence of visible external damage or the Ready for Use indicator unit does not immediately indicate a problem. Customers were instructed to inform all users that if the device is dropped or subjected to severe mechanical shock and the exterior case is still intact, they should immediately perform an operational check as described in the IFU section "Performing the Operational Check" in the Maintenance Chapter. The unit should be taken out of service and Philips Customer Service contacted if the unit is visibly damaged or if the device fails the operational check, i.e., if the RFU indicator changes to a red-X or the device emits a periodic audible chirp, as described in the IFU. Customers were instructed to insert a copy of the customer notice into each copy of the HeartStart MRx IFU. If you need further information or support concerning this notification, please contact your local Philips representative or call the firm at 1-800-722-9377.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.