Philips North America Llc Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS)
Brand
Philips North America Llc
Lot Codes / Batch Numbers
S/N and UDI: 78189 (01)00884838068452, 78272 (01)00884838068452, 78294 (01)00884838068452, 78295 (01)00884838068452, 78332 (01)00884838068452, 78335 (01)00884838068452, 78340 (01)00884838068452, 78345 (01)00884838068452, 78249 (01)00884838068452, 78324 (01)00884838068452, 78326 (01)00884838068452
Products Sold
S/N and UDI: 78189 (01)00884838068452; 78272 (01)00884838068452; 78294 (01)00884838068452; 78295 (01)00884838068452; 78332 (01)00884838068452; 78335 (01)00884838068452; 78340 (01)00884838068452; 78345 (01)00884838068452; 78249 (01)00884838068452; 78324 (01)00884838068452; 78326 (01)00884838068452
Philips North America Llc is recalling Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic devi due to Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
Recommended Action
Per FDA guidance
Philips North America issued U.S. customers letters via certified mail using a 3rd Party service. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall , health risk and action to take: The Philips Ingenia 1.5T, Ingenia 1.5T Evolution, Ingenia 3.0T, and Ingenia Elition X MR systems can continue to fulfill clinical use as normal. However, in case of technical problems related to off cooling of the magnet, magnet pressure or Helium boil-off, please stop use and contact your local Philips service representative immediately and reference FCO78100530. A Philips Field Service Engineer (FSE) will contact you to schedule an on-site visit to resolve the issue by inspecting the potentially affected system(s) to confirm if the non-conforming pressure relief (42 psi) has been installed (reference FCO78100530). If your system has a 42 psi valve, the Phillips FSE will replace the valve with the correct 3.75 psi valve. you need any further information or support concerning this issue, please contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO78100530.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026