Philips Respironics, Inc. Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B
Brand
Philips Respironics, Inc.
Lot Codes / Batch Numbers
Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02
Products Sold
Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02
Philips Respironics, Inc. is recalling Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, due to Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Recommended Action
Per FDA guidance
On July 26, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were advised of the two software issues that the firm had identified related to pressure increase. Customers received guidance on how to identify affected products: Devices with software versions devices with software versions 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01 and 1.06.02 are impacted. Customers were provided with actions to take until the upcoming software fix is implemented on your device. Philips will be releasing a software correction to resolve this issue, and will contact customers when the software correction is released. If you need any further information or support concerning this issue, please contact Philips Customer Care Solutions Center Trilogy Evo: +1-800-345-6443 Option 2, Option 1 (Available 24/7) Trilogy EV300: +1-800-722-9377 (Available 24/7)
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026