Pioneer Surgical Technology Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.
Brand
Pioneer Surgical Technology
Lot Codes / Batch Numbers
Lots 306456, 306457, 306613, 306712, 306713, 306714, 306816, 306817, 306818, 306819, 306820, 306821, 306822, 307526, 307527, 307528, 307667, 307668, 308510, 308938, 310662, 310663, 310664, 310665, 311718, 312169, 519935 and 521042.
Products Sold
Lots 306456, 306457, 306613, 306712, 306713, 306714, 306816, 306817, 306818, 306819, 306820, 306821, 306822, 307526, 307527, 307528, 307667, 307668, 308510, 308938, 310662, 310663, 310664, 310665, 311718, 312169, 519935 and 521042.
Pioneer Surgical Technology is recalling Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP. due to The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.
Recommended Action
Per FDA guidance
Distributors were notified via letter dated July 29, 2005 to retrieve all caps from inventory and point of use locations and to return them to Pioneer.