Quantum Intersegmental Table (Pivotal Health) – Heat Function Risk (2025)
Inadequate battery safety instructions may increase potential device handling risks.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
Brand
Pivotal Health Solutions, Inc.
Lot Codes / Batch Numbers
All serial numbers.
Products Sold
All serial numbers.
Pivotal Health Solutions, Inc. is recalling Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table due to The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to o. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Recommended Action
Per FDA guidance
Pivotal Health Solutions issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 10/9/2025 via certified mail and Fed Ex. The notice explained the issue with the device, risk to health and requested the following: Immediately stop using table(s) with heat. The heat function must be disabled. You have been identified as having been sold a Quantum Table with heat. The serial numbers of the tables impacted are below. If you are unsure if your table has a heat function, please identify the serial number on the Quantum table(s), located on the lower section of table by the power cable inlet, and verify the serial number is on the list. The notice also supplied instructions for disconnecting and disabling the heat function o the Q400 tables. Contact information for questions: o Matt White (matt.white@pivotalhealthsolutions.com) or Julee Driver (julee.driver@pivotalhealthsolutions.com), Quality Dept, at 605-753-0110 o Brenda Anderson (brenda.anderson@pivotalhealthsolutions.com), Service Dept at 605-753-0110
Verify this information on the official source
This page aggregates publicly available data. Always confirm recall details directly with the U.S. Food and Drug Administration (FDA) before taking action.
View official FDA recall noticePage updated: Jan 10, 2026
Important Notice
This page displays information sourced from the U.S. Food and Drug Administration (FDA). IsMyProductRecalled.com aggregates public data for convenience and does not issue recalls. We are not affiliated with any government agency. Always verify recall information with the official source before taking action. This is not medical or legal advice.