Pointe Scientific, Inc. Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s G7517-120,G7517-120, G7517-500, and G7517-1L; 8-G7517-120, 8-G7517-500, 3-G7517-L, 7-G7517-500 and 7-G7517-1000. (Clinical chemistry) Pointe Scientific Liquid Glucose (HEXO) is used for the quantitative determination of glucose in serum. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s G7517-120,G7517-120, G7517-500, and G7517-1L; 8-G7517-120, 8-G7517-500, 3-G7517-L, 7-G7517-500 and 7-G7517-1000. (Clinical chemistry) Pointe Scientific Liquid Glucose (HEXO) is used for the quantitative determination of glucose in serum.
Brand
Pointe Scientific, Inc.
Lot Codes / Batch Numbers
All lots beginning with the numbers 814301, 820502, 823901, 826801, 829401, 814301, 820502, and 831502.
Products Sold
All lots beginning with the numbers 814301, 820502, 823901, 826801, 829401, 814301, 820502, and 831502.
Pointe Scientific, Inc. is recalling Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s due to There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL.
Recommended Action
Per FDA guidance
The firm issued a press release, and an amended press release, on 10/30/09. Consignees were notified by an Urgent: Medical Device Recall letter dated 10/19/09, which identified the problem; and, instructed testing laboratories to inform patients' attending physicians, and determine, with their input, whether confirmation of the previous test results will be required. Customers are to examine their inventory for any of the lot numbers listed on the letter; and, for any product that has been further distributed, those customers need to be identified and notified of this product recall by including a copy of the recall letter. Customers are to complete and return the Recall Return Response form and upon receipt, Pointe Scientific will replace any disposed of reagent. The recall was expanded to include two additional lots by letter dated 11/2/09.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.