Polymedco, Inc POLYMEDCO ULTRA-CRP High Sensitive Assay For the DIMENSION(R) systems, Catalog No. USC300. Contents: 4 x 50 Tests. For In Vitro Diagnostic (IVD) use only. COMMON/USUAL NAME: High-Sensitive C-Reactive Protein --- CLASSIFICATION NAME: C-reactive Protein, Antigen, Antiserum. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
POLYMEDCO ULTRA-CRP High Sensitive Assay For the DIMENSION(R) systems, Catalog No. USC300. Contents: 4 x 50 Tests. For In Vitro Diagnostic (IVD) use only. COMMON/USUAL NAME: High-Sensitive C-Reactive Protein --- CLASSIFICATION NAME: C-reactive Protein, Antigen, Antiserum.
Brand
Polymedco, Inc
Lot Codes / Batch Numbers
Lot numbers CP4274-0769M (EXP. 09/04) and CP4335-0812M (EXP. 11/04)
Products Sold
Lot numbers CP4274-0769M (EXP. 09/04) and CP4335-0812M (EXP. 11/04)
Polymedco, Inc is recalling POLYMEDCO ULTRA-CRP High Sensitive Assay For the DIMENSION(R) systems, Catalog No. USC300. Contents: due to Firm was notified by the foreign manufacturer that certain lots of Ultra-CRP are prone to prozone (a condition of antibody antigen excess) and do not . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Firm was notified by the foreign manufacturer that certain lots of Ultra-CRP are prone to prozone (a condition of antibody antigen excess) and do not meet the labeled specifications.
Recommended Action
Per FDA guidance
By phone on 12/3/03 and by mail on 12/5/03.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026