Qualigen Inc FastPack(R) Total PSA Immunoassay; Catalog #: 25000001 Is a paramagnetic particle immunoassay for the in vitro quantitative determination of prostate-specific antigen (PSA) in human serum and plasma as an aid in the management of patients with prostate cancer. The FastPack(R) Total PSA immunoassay is designed for use with the FastPack (R) System. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FastPack(R) Total PSA Immunoassay; Catalog #: 25000001 Is a paramagnetic particle immunoassay for the in vitro quantitative determination of prostate-specific antigen (PSA) in human serum and plasma as an aid in the management of patients with prostate cancer. The FastPack(R) Total PSA immunoassay is designed for use with the FastPack (R) System.
Brand
Qualigen Inc
Lot Codes / Batch Numbers
Lot Numbers: 0807002-901, 0807002-903, 0807002-904, 0807002-905, 0807002-906, 0807002-907, 0807002-908, and 0807002-909.
Products Sold
Lot Numbers: 0807002-901, 0807002-903, 0807002-904, 0807002-905, 0807002-906, 0807002-907, 0807002-908, and 0807002-909.
Qualigen Inc is recalling FastPack(R) Total PSA Immunoassay; Catalog #: 25000001 Is a paramagnetic particle immunoassay for t due to The recall was initiated after the firm investigation of customer complaints of falsely elevated results determined the product was not meeting produc. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The recall was initiated after the firm investigation of customer complaints of falsely elevated results determined the product was not meeting product performance expectations.
Recommended Action
Per FDA guidance
The recall was initiated with Qualigen faxing a copy of the Product Notification letter, dated 9/29/2008, to all customers who have received any sublot of the affected product . Faxes were sent beginning 9/29/08. The recall letter informed the customers that the affected product was not meeting the firm's expectations for product performance. Customers were instructed to please stop using the affected lots immediately as they have been noted to produce falsely elevated results and they should confirm any result that appears to be inconsistent with "a patient's medical history, clinical examination, and other findings". In addition, they may wish to retest any patients that were run with a different FastPack lot. Beginning on 9/30/08, all customers were contacted by phone to confirm that they received the recall letter. The customers were shipped replacement products at no cost and as part of the product replacement process, they were instructed to return any unused Total PSA FastPacks/Kits of lot 0807002 to Qualigen.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.