Quality Electrodynamics LLC TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453
Brand
Quality Electrodynamics LLC
Lot Codes / Batch Numbers
Serial # 1471 2106 1991
Products Sold
Serial # 1471 2106 1991
Quality Electrodynamics LLC is recalling TxRx 15CH Knee Coil 1.5T QED part number:Q7000016 Siemens Model # 10185453 due to Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Recommended Action
Per FDA guidance
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IN, NC