Radiometer America Inc ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb, FHbF). Also, intended for in vitro testing of samples of expired air for the parameters pO2 and pCO2. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, calcium, chloride, glucose, lactate, total bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb, FHbF). Also, intended for in vitro testing of samples of expired air for the parameters pO2 and pCO2.
Brand
Radiometer America Inc
Lot Codes / Batch Numbers
P/N S1720, 944-024, Lot: WA-04
Products Sold
P/N S1720, 944-024; Lot: WA-04
Radiometer America Inc is recalling ABL7XX Calibration 1 Solution S1720. P/N S1720, 944-024 Intended for in vitro testing of samples o due to The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not refle. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm found that for this particular lot that the barcode, used for entering the characteristics of CAL1 Solution into the analyzer, does not reflect the actual values of the solution. As a consequence, the calibration curves for these parameters and thereby the measured results for patient samples and quality controls will be biased.
Recommended Action
Per FDA guidance
On, 9/2/2008 the firm sent out recall notification letters requesting their customers discard any Calibration 1 Solution (CAL1) S1720 (part number 944-024), lot WA-04, complete the attached return form and fax it to 800-736-0601. The firm will send out replacements for all affected product that the firm shows was shipped to their location. Further questions could be directed to Technical Support at 800-736-0600, option 2.