Radiometer America Inc ABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27
Brand
Radiometer America Inc
Lot Codes / Batch Numbers
There are no specific lot codes. (devices equipped with software version 5.2 to 5.27)
Products Sold
There are no specific lot codes. (devices equipped with software version 5.2 to 5.27)
Radiometer America Inc is recalling ABL800 FLEX Series analyzer with FLEXQ module-equipped with software version 5.2 to 5.27 due to In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
In some situations pre-registered patient data is not reset when leaving the pre-registration mode in the software. This occurs when a Safe Pico blood sampler associated with a specific patient is pre-registered in the analyzer by scanning a barcode.
Recommended Action
Per FDA guidance
Radiometer America issued a recall letter to their U.S. customers dated 1/16/2007. The letter informs the customers of this software anomaly and the conditions underwhich analyzer results can be stored with incorrect Patient Demographics. The letter goes on to inform the customers that a Radiometer Service Representative will be visiting each customer location shortly in order to upgrade the affected software. In the meantime, the letter recommends that the customer(s) not pre-register their samples. The letter is accompanied by an attached FAX FORM which the customer is asked to sign and complete in order to verify receipt of the recall letter and to fax back to the recalling firm.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, CA, CO, IL, KS, KY, MI, MO, NM, NY, NC, OH, SD, TX, UT