Radiometer America Inc Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on ABL800 Flex Series blood gas analyzers.
Brand
Radiometer America Inc
Lot Codes / Batch Numbers
Serial #905-671, 902-754
Products Sold
Serial #905-671, 902-754
Radiometer America Inc is recalling Pump waste tube for blood gas analysis kit. One yellowish/white rubber tube per plastic bag. Used on due to Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrect. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Tube may break. The firm discovered that the yellowish/white tubes used in their blood gas analyzers (ABL800 Flex Series) were manufactured incorrectly. These tubes have the potential to break off at the black plastic connectors. The breakage may occur during the installation of the tubes or after a short time in use.
Recommended Action
Per FDA guidance
By telephone call or service representative visit beginning July 18, 2007, Consumers/End Users are asked to locate the analyzers and yearly service kits concerned and replace all pump tubes with new ones.