Radiometer America Inc SafeClinitubes, Capillary caps (purple), 905-787. The defective capillary caps may be included in capillary tube kits as follows: Clinitubes Kits: 942-890 (D941P-240-85x1), Lot Code: R0048; Clinitubes Kits: 942-892 (D957P-70-100x1), Lot Codes: R0045 through R0050; Clinitubes Kits: 942-893 (D957P-70-125x1), Lot Codes: R0058 through R0063; Clinitubes Kits: 942-898 (D957P-70-70x1), Lot Code: R0019. Each of the kits listed above contains one (1) bag of Capillary Caps, 905-787. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SafeClinitubes, Capillary caps (purple), 905-787. The defective capillary caps may be included in capillary tube kits as follows: Clinitubes Kits: 942-890 (D941P-240-85x1), Lot Code: R0048; Clinitubes Kits: 942-892 (D957P-70-100x1), Lot Codes: R0045 through R0050; Clinitubes Kits: 942-893 (D957P-70-125x1), Lot Codes: R0058 through R0063; Clinitubes Kits: 942-898 (D957P-70-70x1), Lot Code: R0019. Each of the kits listed above contains one (1) bag of Capillary Caps, 905-787.
Brand
Radiometer America Inc
Lot Codes / Batch Numbers
Capillary caps with the Lot #0605068 or Lot #0605069.
Products Sold
Capillary caps with the Lot #0605068 or Lot #0605069.
Radiometer America Inc is recalling SafeClinitubes, Capillary caps (purple), 905-787. The defective capillary caps may be included in c due to Leaking Capillary Caps- The dimensions and shapes of the capillary caps do not fit the capillary tubes correctly. This can cause the capillary tubes. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Leaking Capillary Caps- The dimensions and shapes of the capillary caps do not fit the capillary tubes correctly. This can cause the capillary tubes to leak.
Recommended Action
Per FDA guidance
The firm sent a recall letter to their customers on 6/29/2006. The letter identifies the problem and kit numbers and lot numbers for the affected kits, and as well as the Lot Numbers for the defective 905-787, capillary caps subject to recall. The letter requests that the customers: 1) Examine their inventories for the presence of the affected lot numbered products; 2) Identify and discard any of the recalled bags contained in the kits or ordered separately; and 3) Return the attached FAX FORM indicating the amount of replacement product required and the address to which the replacement product should be mailed. The letter also provides a telephone number to reach the firm''s Technical Support Department should the customer have any questions concerning the recall letter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026