Randox Laboratories CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826
Brand
Randox Laboratories
Lot Codes / Batch Numbers
GTIN: 05055273201765 Batch: 588434
Products Sold
GTIN: 05055273201765 Batch: 588434
Randox Laboratories is recalling CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concent due to The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and w. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
Recommended Action
Per FDA guidance
Randox UK (manufacturer) extended Urgent Medical Device Removal letter to the US distributor via email on 2/1/23. The distributor contacted the customer via email on 2/6/23. Letter states reason for recall, health risk and action to take: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. If you have any questions or concerns, please contact Randox Technical Services.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
OH, PA, WV