RAYSEARCH LABORATORIES AB Sveavagen 9 Stockholm Sweden RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 . Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .
Brand
RAYSEARCH LABORATORIES AB Sveavagen 9 Stockholm Sweden
Lot Codes / Batch Numbers
Software version: RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 2 including some service packs.
Products Sold
Software version: RayStation 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 2 including some service packs.
RAYSEARCH LABORATORIES AB Sveavagen 9 Stockholm Sweden is recalling RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , due to Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.
Recommended Action
Per FDA guidance
On May 6, 2021, Raysearch issued an Urgent Medical Device Correction notice to customers via letter notifying them an issue was found with the voxelwise worst clinical goals in the Robust evaluation module for specified software versions of RayStation. Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. Actions to be taken by the user Voxelwise worst clinical goal in Robust evaluation - The user is advised to be careful when editing clinical goals in the Robust evaluation module and when evaluating the clinical goals for the voxelwise worst distributions. If a clinical goal needs to be edited, make sure to toggle between the Scenario view and Voxelwise min and max views to trigger an update. Contouring issue - The user is advised to always examine ROI geometries in the transversal view and in the reconstructed sagittal, coronal and 3D views to verify that they have the desired size and shape. - To avoid displaying a mix of manually drawn and reconstructed contours in the same 2D view, the user can explicitly convert the ROI geometry to voxel- or mesh representation by using the keep component tool or the deformation tool. This way all rendered contours will be reconstructed and also be identical to what will be exported. All - Educate planning staff and all users about these issues and the workarounds. - Inspect your product and identify all installed units with the above software version number(s). - Confirm you have read and understood this notice by replying to the notification email. Solution The voxelwise worst issue will be resolved in the next version of RayStation, scheduled for market release in May 2021 (subject to market clearance in some markets). The contouring issue was resolved in RayStation 10B. If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers ca
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.