RefleXion X1 (Reflexion Medical) – Software Configuration Risk (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation
Brand
Reflexion Medical, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Model Number: RXM1000 UDI Codes that includes Serial Numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012
Reflexion Medical, Inc. is recalling RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)2 due to Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potenti. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
Recommended Action
Per FDA guidance
On 07/22/2025, the firm sent an "Urgent Medical Device Recall Safety Notification (Correction)" Letter to customer informing them that RefleXion is initiating a Recall (Correction) of the RefleXion X1 System (model RXM1000). An issue has been identified with a configuration setting for the maximum number of fractions that can be delivered within a specified time period. The system should be configured to allow the delivery of up to two (2) fractions, including completion of partial fractions, for each treatment plan within a 12-hour period. As a result of this issue there is a potential to unintentionally administer more than two treatments in a 12-hour period, potentially leading to serious toxicity. Customer are instructed to: 1) Acknowledge receipt of the email by signing and returning the attached Consignee Acknowledgment Form at their earliest convenience. 2) Until the configuration setting is corrected, RefleXion Medical advises users to employ current radiation therapy procedures within the clinic to perform daily checks of treatment session records for radiation therapy patients to prevent delivery of unintended fractions. 3) It is requested that customers post this notification on or near the RefleXion Medical RXM1000 system until this issue can be resolved. For questions or assistance - email support@reflexion.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.