Respironics California Inc Respironics BiPAP Focus Noninvasive Ventilator System. The BiPAP Focus is a non-invasive, pressure support ventilator used to augment the breathing of patients suffering from acute or chronic respiratory insufficiency, respiratory failure or to maintain airway patency and provide respiratory support to patients who experience obstructive sleep apnea. Nasal and full face masks are used with this device. The BiPAP Focus is intended for use in acute, sub-acute and intra-hospital transport settin Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Respironics BiPAP Focus Noninvasive Ventilator System. The BiPAP Focus is a non-invasive, pressure support ventilator used to augment the breathing of patients suffering from acute or chronic respiratory insufficiency, respiratory failure or to maintain airway patency and provide respiratory support to patients who experience obstructive sleep apnea. Nasal and full face masks are used with this device. The BiPAP Focus is intended for use in acute, sub-acute and intra-hospital transport settin
Brand
Respironics California Inc
Lot Codes / Batch Numbers
Model Numbers: 8-400169-00, 7-400076-00, 8-400179-00, 8-400258-00, 8-400259-00, 8-400260-00, 8-400261-00 and 8-400216-00.
Products Sold
Model Numbers: 8-400169-00, 7-400076-00, 8-400179-00, 8-400258-00, 8-400259-00, 8-400260-00, 8-400261-00 and 8-400216-00.
Respironics California Inc is recalling Respironics BiPAP Focus Noninvasive Ventilator System. The BiPAP Focus is a non-invasive, pressure due to Power Supply failures have occurred on some ventilators. A discrepancy between the power supply strain relief clip and the power cord connector may a. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Power Supply failures have occurred on some ventilators. A discrepancy between the power supply strain relief clip and the power cord connector may allow unwanted movement of the cord and the unit's power supply. A disconnection or intermittent electrical connection between the power cord and the unit's power supply during operation or charging can result in transient power surges, which may cau
Recommended Action
Per FDA guidance
Respironics California, Inc. issued an "Attention - Medical Device Field Correction" letter addressed to the Customer/Distributor dated September 22, 2008. The letter informed consignees of the affected device and instructions for removal, device replacement and return of the Business Response form by fax to 1-760-918-0169. Further question may be addressed to Respironics US Customer Service at 1-9800-345-6443 extension 5 or Respironics International Customer Service at 1-724-387-4000.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.