Respironics, Inc. Power Board Kit for use with the PLV-100 Portable Lifecare repairs Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Power Board Kit for use with the PLV-100 Portable Lifecare repairs
Brand
Respironics, Inc.
Lot Codes / Batch Numbers
Model/Catalog number PLV-100. Circuit board part numbers 1034786 version 5 or 1036038 version 3
Products Sold
Model/Catalog number PLV-100. Circuit board part numbers 1034786 version 5 or 1036038 version 3
Respironics, Inc. is recalling Power Board Kit for use with the PLV-100 Portable Lifecare repairs due to Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure.
Recommended Action
Per FDA guidance
The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/8/08. The letter informed the customer of the problem and the need to contact the firm for a replacement. Contact Respironics Customer Service at 1-877-387-331 for assistance.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
IL, IA, MO, NJ, NC, OH, OK, TX