Rita Medical Systems RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649
Brand
Rita Medical Systems
Lot Codes / Batch Numbers
Lot numbers: Y072103-3, Y082003
Products Sold
Lot numbers: Y072103-3, Y082003
Rita Medical Systems is recalling RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649 due to The product displays inaccurate temperature readings, patient skin burns can occur before getting to higher temperature reading.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product displays inaccurate temperature readings, patient skin burns can occur before getting to higher temperature reading.
Recommended Action
Per FDA guidance
On 10/6/03, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected devices and providing instructions on the recall.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026