Roche Diagnostics Corp. ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
All units.
Products Sold
All units.
Roche Diagnostics Corp. is recalling ACCU-CHEK Go Care Kit; Catalog number 04470052001. Kit includes an ACCU-CHEK Go meter. due to The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent u. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards, and held upwards, while the meter is waiting for the sample.
Recommended Action
Per FDA guidance
Recall letters dated 6/7/05 were sent to all customers. Distributors were requested to send a copy of the notification to all customers of test strips used with the product. Users are instructed to never operate the meter without the optics cover in place, and this information will be included in test strip boxes.