Roche Diagnostics Corp. ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - English language version for the blind or visually impaired, Cat. no 2030802. Ref. no. 12030802001 with meter model no. 768. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - English language version for the blind or visually impaired, Cat. no 2030802. Ref. no. 12030802001 with meter model no. 768.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
All units.
Products Sold
All units.
Roche Diagnostics Corp. is recalling ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - English language version due to The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error m. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Recommended Action
Per FDA guidance
A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.