Roche Diagnostics Corp. CoaguChek brand PT Test Strips; U.S. catalog number 3116247 (professional use), 3116239 (patient self test), ex-U.S. catalog numbers 1937642 [packaged 48 strips per box] and 1937634 [packaged 12 strips per box]. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CoaguChek brand PT Test Strips; U.S. catalog number 3116247 (professional use), 3116239 (patient self test), ex-U.S. catalog numbers 1937642 [packaged 48 strips per box] and 1937634 [packaged 12 strips per box].
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
Catalog number 3116247, Lots 591 and 619, Catalog # 1937642, Lot 600, and Catalog # 1937634, Lot 583. Extended to all lots with an expiration date of on or before March 1, 2005.
Products Sold
Catalog number 3116247 , Lots 591 and 619; Catalog # 1937642, Lot 600; and Catalog # 1937634, Lot 583. Extended to all lots with an expiration date of on or before March 1, 2005.
Roche Diagnostics Corp. is recalling CoaguChek brand PT Test Strips; U.S. catalog number 3116247 (professional use), 3116239 (patient sel due to Erroneous patient results: may be falsely elevated or lowered, due to a packaging defect that allows air and moisture to enter the sealed pouch.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Erroneous patient results: may be falsely elevated or lowered, due to a packaging defect that allows air and moisture to enter the sealed pouch.
Recommended Action
Per FDA guidance
Urgent recall letters dated 9/12/03 were sent to customers of 2 lots on 9/12/03 instructing them to cease using the recalled lots and to inform the attending physician of any patients diagnosed using these lots. Distributors were asked to forward the letter to all customers who had received CoaguChek test strips or system care products. A copy of the recall notification was sent to the entire professional customer list reminding them to always inspect the product for seal integrity. The recall was extended to all lots via 9/26/03 press release. Recall letters were sent to all professional and home users on or about 10/1/03 advising them to inspect the package seal, to run each sample twice (with two test strips). Home users were asked to notify their physician of the problem.