Roche Diagnostics Corp. CoaguChek brand PT Test Strips; U.S. Catalog Number 3116247 (48 strip pack-professional use), 3116239 (12 strip pack-patient self test), Non-U.S. Catalog Numbers 11937642190 [packaged 48 strips per box] and 11937634190 [packaged 12 strips per box]. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CoaguChek brand PT Test Strips; U.S. Catalog Number 3116247 (48 strip pack-professional use), 3116239 (12 strip pack-patient self test), Non-U.S. Catalog Numbers 11937642190 [packaged 48 strips per box] and 11937634190 [packaged 12 strips per box].
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
All lots.
Products Sold
All lots.
Roche Diagnostics Corp. is recalling CoaguChek brand PT Test Strips; U.S. Catalog Number 3116247 (48 strip pack-professional use), 311623 due to Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingred. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Erroneous Test Results: Monitor may display 'error' message or report falsely elevated patient results caused by insufficient amounts of active ingredient (thromboplastin) in the test strips.
Recommended Action
Per FDA guidance
A press release was issued by the firm on 10/19/06. The recall letter dated 10/19/06 sent to distributors who sell to home users instructs them to notify their home users and to discard any of product 3116239 in stock. The recall letter dated 10/19/06 to to home users, instructs them to discontinue use of, and to discard, their test strips and to contact their health care professional. The device correction letter dated 10/19/06 to distributors to professional users instructs them to notify customers and provides instructions for duplicate testing in laboratories. The device correction letter dated 10/19/06 to health care professionals asks that they follow-up with their home users to ensure they follow the instructions to cease home use and provides instructions for dupicate testing in laboratories.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026