Roche Diagnostics Corp. Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
All units.
Products Sold
All units.
Roche Diagnostics Corp. is recalling Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001. due to If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
Recommended Action
Per FDA guidance
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026