Roche Diagnostics Corp. Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
Lot 23717132, exp 5/31/2005.
Products Sold
Lot 23717132; exp 5/31/2005.
Roche Diagnostics Corp. is recalling Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138. due to Coating defect may result is falsely decreased patient and/or control results.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Coating defect may result is falsely decreased patient and/or control results.
Recommended Action
Per FDA guidance
Recall letters dated 7/14/04 were sent to each customer asking that use of these lots be stopped immediately. Testing labs were asked to notify the attending physician for any patients tested with these test strips.