Roche Diagnostics Corp. Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03253678122.
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
Lots 175646 and 174558.
Products Sold
Lots 175646 and 174558.
Roche Diagnostics Corp. is recalling Roche Elecsys Folate II for use on Elecsys 210 Analyzers; Catalog No. 3253678 (Roche Material No. 03 due to Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly hi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Gel-like particle formation; may result in a clot settling in the analyzer measuring cell, which will create a low signal and may cause incorrectly high or low QC and/or patient results depending on the assay method.
Recommended Action
Per FDA guidance
Consignees were informed via letter dated 9/15/06 to perform Folate assays in batch mode and to run controls and liquid flow cleaning upon completion of each batch run. Distributors were requested to notify their customers.