Roche Diagnostics Corp. Roche MagNA Pure LC System. An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions. Catalog numbers 12236931001 and 03670325001 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Roche MagNA Pure LC System. An automated, general purpose instrument for isolation and purification of nucleic acids and assembly of PCR reactions. Catalog numbers 12236931001 and 03670325001
Brand
Roche Diagnostics Corp.
Lot Codes / Batch Numbers
All units.
Products Sold
All units.
Roche Diagnostics Corp. is recalling Roche MagNA Pure LC System. An automated, general purpose instrument for isolation and purification due to The firm has received complaints of reagent splattering on the deck of the instrument and possible cross contamination of samples purified on the inst. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The firm has received complaints of reagent splattering on the deck of the instrument and possible cross contamination of samples purified on the instrument.
Recommended Action
Per FDA guidance
Consignees were notified by letter dated 4/4/06 to notify customers of the problem and provide a work-around.